User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Keithley 177:Microvolt DMM INSTRUMENT: Keithley 177:Microvolt DMM (DTB# 9-18) DATE: 02-Oct-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 31 NUMBER OF LINES: 79 CONFIGURATION: Fluke 5700A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- F W 1.002 ASK+ K X 1.003 HEAD * INITIAL CONDITIONS * 1.004 DISP Allow the UUT a 1 hour warm-up. 1.005 DISP [32] Connect the 5700A and the UUT as follows: 1.005 DISP [32] 2-WIRE COMP 1.005 DISP [32] 5700A TO 177 1.005 DISP [32] OUTPUT-HI ------------> INPUT-HI 1.005 DISP [32] OUTPUT-LO ------------> INPUT-LO 1.005 DISP [32] SENSE-HI ------------> INPUT-HI 1.005 DISP [32] SENSE-LO ------------> INPUT-LO 1.006 HEAD {* DC VOLTS TEST *} 1.007 5700 2 1.0000V 0.0004U 2W 2.001 5700 2 -1.0000V 0.0004U 2W 3.001 5700 20 10.000V 0.004U 2W 4.001 5700 20 -10.000V 0.004U 2W 5.001 5700 200 100.00V 0.04U 2W 6.001 5700 200 -100.00V 0.04U 2W 7.001 5700 1000 1000.0V 0.5U 2W 8.001 5700 1000 -1000.0V 0.5U 2W 9.001 DISP Set UUT to 20mV range. 9.002 ASK- R 9.003 5700 0mV S 2W 9.004 DRAW 8 78 8 78 100 5 66 70 9.005 DISP Adjust UUT front panel ZERO for a display indication 9.005 DISP of 0.000mV +/-0.001mV. [DRAW66,10,F] 9.006 5700 20 19.000mV 0.013U 2W 10.001 5700 20 -19.000mV 0.013U 2W 11.001 ASK+ R 11.002 5700 200 190.00mV 0.11U 2W 12.001 5700 200 -190.00mV 0.11U 2W 13.001 HEAD { * AC VOLTS TEST *} 13.002 5700 200 100.00mV 0.65U 1kH 2W 14.001 5700 2 1.0000V 0.0065U 1kH 2W 15.001 5700 20 10.000V 0.065U 1kH 2W 16.001 5700 200 100.00V 0.65U 1kH 2W 17.001 5700 1000 1000.0V 6.5U 1kH 2W 18.001 5700 20 10.000V 0.085U 45H 2W 19.001 5700 20 10.000V 0.085U 20kH 2W 20.001 HEAD {* DC CURRENT TEST *} 20.002 5700 200 100.00uA 0.21U 2W 21.001 5700 2 1.0000mA 0.0021U 2W 22.001 5700 20 10.000mA 0.021U 2W 23.001 5700 200 100.00mA 0.21U 2W 24.001 5700 2000 1000.0mA 2.1U 2W 25.001 HEAD {* RESISTANCE TEST *} 25.002 ASK- R 25.003 DISP Set UUT to the 200 Ohm range. 25.004 5700 0Z S CW 25.005 DRAW 8 78 8 78 100 5 66 70 25.006 DISP Adjust UUT front panel ZERO for a display indication 25.006 DISP of 00.00 Ohm +/- 0.01 Ohm. [DRAW66,10,F] 25.007 5700 200 190.00Z 0.14U CW 26.001 ASK+ R 26.002 5700 2 1.9000kZ 0.0011U CW 27.001 5700 20 19.000kZ 0.011U CW 28.001 5700 200 190.00kZ 0.11U 2W 29.001 5700 2000 1900.0kZ 1.1U 2W #30.001 JMPT 31.001 Keithley 177:Microvolt DMM (DTB# 9-18) 30.001 5700 20 10.000MZ 0.100U 2W #31.001 5700 20 10.000MZ 0.100U 2W 31.003 END